|
115:412/508 Protein
Determination
spring 2002 Before continuing with today’s material,
I want to say a little about the quiz last time. I’m returning them to those who put down something of substance
– the rest will not want to be embarrassed further. The point of the quiz of course was to bring it home to you that
for this course you need to know the properties of the amino acids,
because they to a first approximation comprise those of proteins. The amino acids with side chains that ionize
are 7: aspartic and glutamic acids, arginine, histidine and lysine,
tyrosine and cysteine. Arginine
is always protonated, since its pKa
is about 12.5, and lysine is almost always protonated, since the “normal”
pKa is about 10, but in special cases the pKa can be as low as 8 and the unprotonated
form be significant. Similarly,
aspartic and glutamic acids are normally deprotonated, with a pKa of 4, but not uncommonly the pKa is higher and the protonated form is significant.
The phenolic OH of tyrosine has a pKa ≈ 10, but rarely ionizes except when
the protein is denatured in base, because tyrosine is otherwise a hydrophobic
amino acid, likely to be buried in the interior of the protein – there
are ways to estimate the % exposure of tyrosines. But the OH may well be involved in a hydrogen
bond. Histidine, with a pKa in proteins of about 7, and cysteine, with
an average pKa for the side chain of perhaps 8.5, are
most likely to have both protonated and unprotonated states involved
in enzyme mechanisms. Of course a-amino groups and carboxyl groups
also ionize, but there is only one of each per polypeptide chain – the
others are in peptide linkage, non-ionizing but able to form hydrogen
bonds which are the basis of secondary structure. The remaining amino acids comprise those
with polar but non-ionizing side chains – serine and threonine deprotonate
only above pH 13, and asparagine and glutamine protonate the amide only
below pH 1 – and the purely hydrophobic, all the others unless you make
glycine a special class because it has no side chain at all.
The polar but non-ionizing do participate significantly in hydrogen
bonds, as do those with ionizing side chains, and potentially the indole
nitrogen of tryptophan, which also protonates only below pH 1. For the rest, remember that CH2 and CH3 groups, and aromatic side chains, are as
hydrophobic as salad oil and benzene.
And never draw structures with five bonds to carbon! The next few lectures are concerned with
protein purification. You may
want to do this to study the protein physically (what is its size, shape,
3-D structure, etc.) or chemically (what is its amino acid sequence,
post-translational modification, etc.), both of which types of study
require that the total protein present is substantially the one you
want to study - at least 95% of the total.
If you want to clone the gene or cDNA for the protein, you want
either to determine at least some of the amino acid sequence, in order
to design oligonucleotide probes for the cDNA sequence, or to make antibodies
to the protein, to look for clones expressing it.
In the latter case you may not have to purify it completely,
because you can also select a monoclonal antibody for it; but this is
work of another kind. Study of its biological activity requires
less purification, because you have only to remove other proteins which
interfere with this study, but the only way to be sure that one is free
of such interference is to remove all other proteins.
Nowadays, you may want to purify the protein for therapeutical
purposes, and need to have it very pure indeed.
The usual way to determine if the protein is pure is gel electrophoresis,
which will be covered much later. In purifying a protein, we are trying to
keep it, specifically, and throw away all other proteins, until we have
only the protein we want. To measure whether progress is being made
in purification you need to measure the 'specific
activity', the ratio of the amount of the protein you are interested
in, as measured by some specific
assay, to the amount of all
protein present, as measured by a general assay of all protein present. You determine the efficiency of a given purification
step by how much the specific activity is increased after it. And sometimes you need to measure the protein
concentration because it is critical for the success of a given step
- for instance, (NH4)2SO4 precipitation won't work if the solution
is too dilute. On the other
hand, there are times when specific activity is less important than
total activity. For instance, if you want to test whether a
protein is being secreted into culture medium at a higher level, you
care about total activity in the medium; increased specific activity
might just mean that less other
protein was being secreted. Protein determination We should
therefore discuss how we measure protein in general. Most of you have probably done this at some time, but it is worth
discussing various assays, since which you use may depend on a particular
situation. A protein assay should have the following
characteristics, usually but not always in about this order of importance:
it should be sensitive (i.e. measuring low levels of protein), quick,
specific (i.e. not measuring other compounds present, nor affected negatively
by their presence), non-destructive, and invariant from protein to protein
(i.e. giving the same results per mg of protein present).
No one assay is perfect with respect to all considerations -
else it would have displaced all others - but which you use may depend
on which of the above considerations is most important at the time. Note also that almost all assays do give different results with
different proteins, because they are affected by the amino acid composition
of the protein, and therefore cannot be said to be completely accurate
unless they have been standardized using a solution of the protein you
are interested in, whose concentration has been determined by the dry
weight of protein per ml. And
even that is not accurate if the protein contains a significant amount
of non-protein material such as carbohydrate.
As a standard for the method used during purification one usually
uses bovine serum albumin, which is cheap and average in composition;
one prepares a standard curve, measuring the absorbance of known amounts.
However, it gives more color in the Coomassie Blue assay than
the average of other proteins. For
best accuracy one should dry down a measured volume of the standard
solution to obtain the actual weight concentration. To comment on those on the sheets I am handing
out, and a few others: the biuret
reaction, which measures the absorbance at 540 nm of Cu++
bound to peptide bonds, is the most invariant from protein to protein
- because it is a reaction of the peptide bond, and therefore unaffected
by amino acid composition, though the weight
of protein will be affected by composition, polytyrosine obviously
weighing more per peptide bond than polyglycine.
It is also very insensitive, requiring at least 1 mg, and destructive,
using strong base; it is therefore rarely used nowadays. However, it is more sensitive in the ultraviolet; the methods of
Goa and Itzhaki & Gill take advantage of the fact that the absorbance
of the biuret complex is greater in the near UV, 310 or 320 nm; but
it's not a peak. A variation of the biuret reaction is that
of Klungsøyr. Cu++ complexed with protein in 0.1
m NaOH moves with it through
a little Sephadex column, while free Cu++ is adsorbed at
the top of the column. One can
then measure the Cu++ carried through as its complex with
diethyldithiocarbamate . This is as sensitive as the Lowry method -
0.05 to 0.4 mg - but as invariant as the biuret method. But it is not adapted to doing many samples
routinely, since each must be chromatographed. (The Sephadex spin column methodology much used in molecular biology
for small samples would be useful here.) It can be useful to give a fairly absolute measurement of the concentration
of a pure protein, which can be translated into a molar concentration
when you are chemically modifying the protein and want a ratio of moles
modifier/mole protein. Measuring
protein by a method depending on reaction with the side chains may be
inappropriate if you are chemically modifying the side chains which
react in the method. The most frequently referred to paper in
biochemistry - and in all science - is that by Lowry, Rosebrough, Farr and Randall, describing what is also called
the phenol or Folin method, which uses a combination of the biuret absorbance
of Cu++ bound to the peptide bond and Cu+ produced
by reduction of the Cu++ by tyrosine, cysteine and tryptophan;
the Cu+ is measured by its reduction of arsenomolybdate,
as in the Nelson-Somogyi method for reducing sugars and the Fiske-SubbaRow
method for inorganic phosphate. This
method is fairly sensitive, measuring 0.05 - 0.4 mg protein, and only
moderately variable from protein to protein; but a wide variety of compounds
which may be present in a protein solution interfere, notably SH compounds
such as 2-mercaptoethanol and compounds with the HNCH2CH2O- structure such as triethanolamine and
morpholine buffers. The interference
can generally be dealt with, as by first precipitating the protein with
a combination of trichloroacetic acid (TCA) and deoxycholate, but this
make even slower an already relatively slow procedure, requiring 40
minutes' incubation besides pipetting time.
It is also not strictly linear, though if only absorbances less
than 0.4 are used this is not a problem. A related procedure has been ballyhooed
by Pierce Chemical Co. It is
based on the chelation of Cu+ by bicinchoninic
acid and the absorption of this complex.
The method is apparently somewhat less affected by interfering
compounds than the Lowry method, but it seems to me that any compound
capable of reducing Cu++ to Cu+ would still interfere
strongly. The color development is probably faster than
in the Lowry procedure, but generally incomplete; sensitivity is increased
by incubating at a temperature such as 60°.
This would tend to cause variability from assay to assay and
require a standard curve each time. The main current competition of the Lowry
method, probably now much more used, is the 'dye-binding' or Coomassie Blue or Bradford method. BioRad sells the reagent made up, but unless
money is no object it is cheaper to make up your own; however, a standard
curve should be run regularly. The
method depends on the fact that a solution of the dye Coomassie Blue
G-250 in acid - phosphoric or perchloric, pH about 1 - is orange, lmax 466 nm, but protein displaces two protons
from the dye, generating a blue form which is measured at 595 nm. There is a good deal of background absorption
by the reagent, which can cause variation in measurements, particularly
if poorly matched cuvettes are used; usually one is adding a small amount
of protein to a large amount of the reagent, so that errors due to variation
in pipetting the reagent are small.
The method is perhaps 4x more sensitive than the Lowry method,
and much less affected by interfering substances, except for one important
one: sodium dodecyl sulfate (SDS), which binds to and favors formation
of a neutral green form, lmax = 650 nm. It should be remembered that any compound present in high enough
concentration to shift the pH into the range where the dye is blue anyway
will also interfere; a large amount of a high concentration of (NH4)2SO4
could for instance do this. The
method is moderately variable from protein to protein; the paper by
Compton and Jones showed that binding is mainly to arginine groups,
to a lesser degree hydrophobic. It can be non-linear at very low protein concentration;
the paper by Zor & Selinger shows that this is due to equilibration
between the three ionic forms, and can be linearized and made more sensitive
by reading both A466
and A595
and plotting
A595/A466. This is more trouble,
unless you have a spectrophotometer which reads at two wavelengths and
give you the ratio automatically; but it expands the useful concentration
range both down and up, and makes it more sensitive because the A595/A466 ratio changes more with protein concentration than the A595 alone.
The other disadvantage is that the reagent tends to stick to
cuvettes, particularly good quartz cuvettes.
If you can't afford a single use of disposable cuvettes, you
can remove the blue stain with acetone (not for plastic cuvettes!),
0.1 m NaOH, or SDS. The most important other method is ultraviolet absorbance. Absorbance at 280 nm is quite variable from
protein to protein (depending on the content of tyrosine and tryptophan),
subject to interference in very crude extracts (though nucleic acid
content can be corrected for by A260),
and only middling sensitive (an average protein has an E280 of about 1 for a 1 mg/ml solution). But the method is the quickest possible (pipet
it into the cuvette and read) and non-destructive ; it is therefore
the best method when you have lots of fractions to read, as from column
chromatography, and don't care too much about the exact concentration,
only where there's protein and where there isn't. UV absorbance can also be read in the far
ultraviolet, typically at 205 nm, where the peptide bond absorbs and
a 1 mg/ml solution has an absorbance of about 30; the variation depends
mostly on the tyr + tryp content and thus the extinction coefficient
can be corrected for this, E205 = 27.0 + 120(A280/A205). Some far-UV monitors such as the LKB Uvicord read at 206
nm. The problem here is that
nearly everything else, except SO4= and
ClO4-, also absorbs at this wavelength, and
you have to have the best available quartz cuvettes, very clean, and
a good deuterium lamp to read at this wavelength.
Since the sample usually must be diluted to decrease absorbance
by the buffer, the increase in sensitivity is not fully realized. Where this method, and one other I'll mention,
is most useful is in column or HPLC chromatography of peptides from
enzymatic or chemical cleavage, being separated for sequence determination.
Such peptides, particularly small ones, may not contain tyrosine
or tryptophan, and thus you want to monitor fractions by a method not
depending on the amino acid composition. (The biuret method of course
is too insensitive.) The other method for this situation is basic
hydrolysis (to free amino acids) followed by quantitative ninhydrin reaction.
This is quite sensitive and insensitive to amino acid composition
except for proline (yellow product rather than purple), but of course
it's a lot of trouble for a large number of fractions. Look for it in Methods in Enzymology vol. 11, paper by C.H.W. Hirs. Another reagent which made a splash for
a while is fluorescamine,
a reagent developed at Hoffman-LaRoche which gives a fluorescent product
with primary amines. This can
be very sensitive, but of course cannot be used in amine buffers such
as Tris, and in my experience fluorescent impurities are a problem.
It and similar reagents such as o-phthalaldehyde
find their best use as replacements for ninhydrin in measuring free
amino acids in amino acid
analysis (the determination of the amino acid composition of a protein),
which is really the most absolute determination of protein concentration
- you can add up the amounts of all the amino acids present and calculate
the protein concentration - but too expensive and equipment-intensive
for routine use. Where amino
acid analysis is important is in measuring chemical modification of
the protein; it is generally easy to measure the amount of modification,
more difficult to measure the molar concentration of the protein exactly
in order to determine moles modification/mole protein. A couple of older methods should be mentioned.
Turbidimetry - the optical density (not absorbance) of a cloudy suspension
of protein denatured with TCA or other acids - used to be used, but
hardly ever is since the development of the Lowry method. It is linear over only a fairly narrow range,
dependent on very careful control of conditions, and not very sensitive.
I probably mention it only because we did it in a lab when I
was a graduate student. Kjeldahl
nitrogen determination - reduction of essentially all nitrogen in a
sample to ammonia, which is distilled and measured by titration or the
Nessler reaction - is ancient and highly non-specific, but useful under
one condition (not likely to be of interest to people in this course):
when the sample is highly insoluble, as for protein content of seeds.
One could certainly use basic hydrolysis and ninhydrin here. The article by Stoschak in Methods in Enzymology
vol. 182 includes a method she devised using colloidal gold. I have no experience with this; it seems to
be extremely sensitive but also dependent on careful control of conditions. To summarize: one normally uses the Lowry
or Coomassie Blue method on each defined stage of a purification, though
one probably uses UV absorbance on column chromatographic fractions. But it is good to know a lot of possible methods,
because you may have reason to use another under some specific condition,
such as when you need to know the molar concentration of your protein,
or when you have modified amino acids that react in your usual method. |